Show news: from last 30 days (default), 60 days, 90 days, last visit or all

Syringe Driver Survey Database migration to MedicinesComplete - January 2023

Syringe Driver Survey Database has moved to MedicinesComplete and is no longer available on palliativedrugs.com.

It has been renamed to PCF’s Syringe Driver Database and is now accessed through Drug Compatibility Checker.

Drug Compatibility Checker, through MedicinesComplete is a new tool providing essential compatibility knowledge to support the administration of injectable drugs combining published data and clinical practice reports.

Find out more about Drug Compatibility Checker:https://about.medicinescomplete.com/publication/drug-compatibility-checker/

If you have a subscription to Palliative Care Formulary through MedicinesComplete, you will have access to PCF’s Syringe Driver Database through Drug Compatibility Checker, at no charge through 2023. 

To contribute to PCF’s Syringe Driver Database please continue to submit a clinical practice report using palliativedrugs.com here: https://www.palliativedrugs.com/syringe-driver-database.html

To learn more about a subscription to the Palliative Care Formulary through MedicinesComplete, please contact us here: https://info.medicinescomplete.com/sales

If you have any questions regarding access to PCF’s Syringe Driver Database or would like to renew your subscription to Palliative Care Formulary, please contact pharmpress-support@rpharms.com.


CME T34 syringe pumps now require foam pads to be added to the battery compartment

3rd April 2019

All Caesarea Medical Electronics (CME) T34 Ambulatory syringe pumps now need a foam pad to be added to the battery compartment to ensure the battery rests securely against the contacts. Kits should be ordered from CME to do this.

This information is in the latest CME field safety notice (FSN2018-001_Updated) and follows the previous alerts in March and September 2018, highlighting the possible loss of function due to inadequate battery connection.

Click here to view